balloon pump should be 2-3 cm away from subclavian artery
- related: cardiac supportive devices
- tags: #permanent
- too high: occlude CNS vessels
- too low: occlude abdominal vessels
either case can lead to organ ischemia
Balloon should be located just below left subclavian artery, at the level of AP window on CXR.1