sensitivity of pacemakers are different for atrium and ventricle
- related: cardiac pacemaker
- tags: #permanent
- atrium usually need 0.5 mV sensing
- ventricles need 2.0 mV sensing
- decreasing sensitivity leads to more false positives
- increasing sensing makes it harder to detect native rhythm and leads to obligatory pacing
- when in doubt, leave the sensing alone