sensitivity of pacemakers are different for atrium and ventricle


  • atrium usually need 0.5 mV sensing
  • ventricles need 2.0 mV sensing
  • decreasing sensitivity leads to more false positives
  • increasing sensing makes it harder to detect native rhythm and leads to obligatory pacing
  • when in doubt, leave the sensing alone